Introduction

The nervous system consists of transcriptomically distinct neuronal cell types that reflect differences in sensing capabilities, connectivity, and function. Mapping their respective functions represents one of the major goals and challenges for modern neuroscience1. To this end, optogenetics has facilitated the untangling of neural networks by using light-activated, genetically encoded opsins to selectively manipulate the activity of distinct neuronal cell types with spatial precision2. However, this functionality has been limited to the brain due to constraints associated with peripheral light delivery: body tissues typically lack a stable interface for securing fiber optics and the inflexible nature of these optical probes would cause shearing of tissues and nerves during an animal’s natural movements. As a result, a cell-type-specific understanding of the peripheral nervous system in freely behaving animals is severely lacking.

This is exemplified by the vagus nerve, which provides the only direct neural communication between internal organs and the brain. Peripheral endings of vagal afferent fibers respond to a broad array of stimuli, including hormones, osmolytes, changes in pH, and mechanical distention that have diverging functions and contributions to behavior3. All of its diverse sensory cell bodies reside together within the nodose ganglia3, but conventional viral and transgenic methods for targeting genetically distinct neuronal populations do not permit organ-specific manipulations. Although pioneering studies have used fiber optics to optogenetically manipulate mouse vagal afferents with organ specificity, these studies were conducted under anesthesia to investigate autonomic functions4,5. Studying functions beyond reflexes, such as gastrointestinal mechanisms of satiation, requires a more flexible approach. Given the widespread interest in using vagal nerve stimulation for treating obesity and other neurological disorders6,7, a key priority for this research field is to attain cell-type- and organ-specific manipulations of the vagus nerve in animals that are awake. Accordingly, we set out to develop a biocompatible, wireless optogenetic device for organ-specific light delivery.

Advances in wireless technologies have enabled the internalization of light sources, bypassing physical constraints associated with fiber-optic cables, and driving a shift in what is possible with optogenetics. Wireless, radio-frequency (RF)-powered devices were miniaturized for microscale light-emitting diode (µLED) insertion into the brain8 and recently encased within stretchable and impermeable tethers for securing them onto subdermal tissues9. Despite these developments, organ-restricted illumination remains a challenge. A wirelessly powered µLED that is secured to the rat bladder using a circumferential elastomer sleeve could enable a similar level of functionality10, but this approach impedes organ expansion. Efforts to wirelessly manipulate neural organ function in awake mice include studies that sutured an µLED onto the heart surface for pacemaking11 or intestine surface for controlling colonic motility12. However, these devices were not described as being functional for >8 days10,11,12, a limitation for conducting behavioral studies given the extended recovery periods required after thoracic and abdominal device implantation. Moreover, affixing the µLED to the target organ surface results in light back-scatter and nonspecific optogenetic illumination of nearby tissues. No device has yet enabled chronic and durable cell-type-specific optogenetic manipulation of peripheral neurons inside of an organ.

Here, we describe the development of a durable, multimodal, wireless platform that enables optogenetic stimulation of peripheral neurons within organs in a high-performance manner. The miniaturized wireless device is fully implantable, utilizing a soft, thin, and low-modulus tether for targeting a µLED inside an organ. A unique fabrication method is employed to make a robust, µLED-housing tether, permitting long-term (>1 month), intimate interfacing with peripheral nerve endings in freely behaving mice. These optogenetic implants can selectively and independently manipulate peripheral nerve activity within multiple target organs in the same animal using a monolithic design. In addition, a channel isolation strategy is introduced for powering multiple cages using a single RF transmitter. Coupled with an advanced coil-antenna approach, a single telemetry system provides reliable wireless power in eight individual homecages, overcoming cage limitations of other wireless and fiber-optic-based systems. Precise targeting of a µLED within the stomach revealed an unexpected role for putative, gastric chemosensors in suppressing appetite and revealed a valence mechanism by which appetite suppression occurs.

Results

Organ-specific, wireless gastric optogenetic device

An illustration of the fully implantable wireless device shows the general strategy for targeting a µLED inside the stomach (Fig. 1a, b). The device consists of an analog, front-end electronic circuit for RF harvesting (5.5 mm radius and 1 mm thickness) and a tether that supplies current to a µLED. It harvests RF energy from a remotely located wireless RF-power system, converts RF energy into optical energy, and illuminates targeted regions in the stomach. The µLED is situated in the middle rather than the end of a tether, allowing the tether to be threaded in and out of the stomach and secured at two contact points. We found that the tether remains secure with purse-string sutures. The ultra-thin tether (0.4 mm wide by 0.2 mm thick) is more than three times smaller than insulin syringe needles used for intraperitoneal injections and tubing used for intragastric infusions13.

Fig. 1: Development of a soft, wireless gastric optogenetic implant with a pre-curved, sandwiched tether.
figure 1

a Illustration of a soft, wireless gastric optogenetic implant with a pre-curved, sandwiched tether. b Image of wireless LED operation in the stomach; scale bar 5 mm. c Procedures for device fabrication; scale bar 5 mm. d Three-dimensional modeling of the mechanics for the pre-curved, sandwiched structure. e Plots of output power vs. curvatures of a tether. The legend numbers represent the curvature of the device as the length of the arc for a radius of 2 mm (see Supplementary Fig. 3 for equation); smaller numbers represent sharper U-shape angle. f Measurement results of device lifetime cycling test for both structures when strain applied in the horizontal (top) and vertical direction (bottom). g Measurements of device lifetime for the pre- and post-curved structure when implanted (pre-curved, n = 8; post-curved, n = 8). Bar graphs are mean ± SEM. Statistical comparison was made using two-tailed t test; ***p < 0.001.

Essential features that allow for long-lasting operation of the ultra-thin tether are a pre-curved, sandwiched construction. In our prototype, the harvester and µLED were connected with thin copper (12 μm) electrical interconnects on top of flexible and durable polyimide (18 μm) substrate, and then coated with a biocompatible silicone polymer, polydimethylsiloxane (PDMS). However, this design exhibited poor durability and post hoc analysis revealed µLED tether damage likely caused by a mechanical strain. To increase durability, the µLED was sandwiched in between a second copper/polyimide bilayer, which also provided additional electrical contact (Fig. 1c; assembly steps 1 and 2). We further postulated that coating the tether with silicone in a curved position (pre-curved) would decrease strain compared to a tether that was coated in a flat orientation and then bent when securing it inside the stomach (post-curved). This was achieved by suspending the tether in a bent position, pipetting small amounts of melted silicone around the µLED, and coating the remaining components using a simple dipping process (Fig. 1c; assembly step 3). This resulted in a thin, soft, and lightweight (~380 mg), wireless, gastric optogenetic implant (Fig. 1c; assembly step 4). The compliant, low-modulus properties eliminated constraints on the natural motions of the animal while also minimizing mechanical strain at the connecting joints.

Three-dimensional (3D) modeling of the mechanics showed that the maximum strain in the copper traces and PDMS coating of the pre-curved tether (<800 Pa) was dramatically reduced compared to the post-curved tether strain (<6600 Pa) (Fig. 1d and Supplementary Fig. 2). We further optimized the tether by mechanically testing various curvatures and identified a pre-curved configuration that was functional beyond 200 kilocycles (kc) (Fig. 1e and Supplementary Fig. 3). Lifetime mechanical cycle tests with a significant load (0.03 kgF) revealed that the pre-curved structure with a radius of 1.15 mm was functional for 200 kc, a nearly 10-fold improvement compared to the post-curved structure (Fig. 1f and Supplementary Fig. 4). Although there was improved durability with pre-curved structures that had a radius of 2.87 and 0.72 mm, they were not as durable as 1.15 mm, likely because 2.87 mm is too similar to the flat structure, whereas the sharp angle with a radius of 0.72 mm interferes with µLED contact with the pad. The device was also subjected to waterproof testing by submerging into a heated saline solution, revealing that it remained continually functional for over 2 months, even in extreme temperatures (Supplementary Fig. 5). Heat dissipation is another factor that can limit device functionality since nerve endings in the gastrointestinal tract can be temperature sensitive14. Thermal assessment of the wireless optogenetic implant demonstrated minimal temperature increases (~0.2 °C; Supplementary Fig. 6) during typical operating conditions (10 and 20 Hz with 5 ms light pulse; 10 and 5% duty cycles). Consistent with this, calculation of specific absorption rate (SAR) using a finite-element method analysis tool showed that the SAR distribution against localized RF exposure is below IEEE guidelines15 (Supplementary Fig. 7). Finally, tests in mice showed that the pre-curved, sandwiched tether was functional for over a month, while the post-curved structure stopped working 3 days after implantation (Fig. 1g).

Efficient and high-throughput, wireless optoelectronic systems

The practical use of optogenetics depends on reliable and cost-effective light delivery in multiple animal subjects. A complete laser-based, optogenetics setup remains cost-prohibitive for many labs, given that each animal subject requires a laser, fiber-optic cannula, fiber-optic patch cord, and rotary joint to decrease physical constraints of a patch cord16. Wireless optogenetics is similarly limited, typically requiring a single RF-power generator for each homecage17. Multiple RF-power generators can be used, but they must be operated at least 1 m apart from each other to avoid electromagnetic interference18 (Supplementary Fig. 8). Together, these constraints limit the group sizes used for studies, restrict the duration and type of behavioral experiments that can be conducted, and overall prohibit the high-throughput utilization of optogenetics.

To overcome these limitations, we developed a multiplex approach to power eight individual cages with a single RF-power generator. The wireless telemetry system consists of an RF-power generator, controller, RF multiplexer, decoupling multiplexer, and an antenna set for each of the 8 cages (Fig. 2a); each antenna set is made of a pair of top and bottom coil structure. Simultaneous and independent control of the eight cages is achieved with coupling and decoupling circuits that manage the tuning of antennas to operational (13.56 MHz) and non-operational (100 MHz) device powering frequencies. For example, when the controller selects antenna set 6, the RF multiplexer tunes antenna set 6 to 13.56 MHz and decoupling multiplexer detunes the other antenna sets to 100 MHz (Fig. 2b). The other antennas that are detuned and deactivated can only pass a negligible amount of energy at a frequency of 100 MHz, which significantly deviates from the resonant frequency of 13.56 MHz. Therefore, the other seven antenna sets do not cause interference even when directly adjacent to the actuating 13.56 MHz antenna; this was confirmed with electromagnetic simulation results and validation experiments in vivo (Fig. 2c, d, Supplementary Movies 1, 2, and Supplementary Fig. 9). Since optogenetics typically requires brief intermittent light pulses to avoid depolarization block16, this strategy can be used to toggle between multiple cages to deliver intermittent light pulses. Therefore, the limiting factors for the number of cages that can be operated simultaneously with a single RF-power generator are the stimulation frequency and duration of light pulses. With the proposed arrangement (Fig. 2a), we conducted experiments within eight cages simultaneously using 20 Hz and 5 ms pulse duration stimulation parameters. This extends the high-throughput utilizing optogenetics to do the experiments with at least eight mice (supposed a mouse in a cage), a typical group size, at that same time. For example, measurements of food intake require 4 h for each animal in the group. To complete the analysis of feeding behavior for two groups of animals (experimental and control), each of which has eight animals, it only takes 8 h, while approaches using existing wireless TX system, a single power source coupled with a single cage, demands 64 h (8 × 8 h). This makes it less ideal for longitudinal experiments, in particular those required for most obesity experimental designs, where a device needs to be chronically implanted for >2 months. Also, through the modification of the multiplexer board and controller, simultaneous activation of 16 cages at 10 Hz or 32 cages at 5 Hz frequency is possible when using a 5 ms light pulse duration.

Fig. 2: Electrical characteristics of multiple cage wireless power TX system and multiple organ stimulation devices.
figure 2

a Schematic illustration of the proposed wireless power TX system for high-throughput phenotyping of neural pathways. b Functional block diagram of the proposed wireless power TX system. c Electromagnetic simulation of wireless coverage for the proposed TX system; scale bar 10 cm. d Picture of the TX system; scale bar 10 cm. e Representative magnetic field distributions in antenna set 6 (top) and comparisons of wireless coverage for the proposed system and other wireless power TX systems (bottom); scale bar 10 cm. f Illustration of wireless operation of a scalable, multimodal wireless gastric optogenetic implant (left), images of an animal with the device implanted (middle), and image of the device (right); scale bar 1 cm. MUX; multiplexer, Decou.; decoupling, SW; switch, Cap; capacitor, Mat. B.; RF matching board, Ant.; antenna set.

In addition to couplings, wireless coverage remains a significant limitation for optogenetic experiments. Conventional systems utilize a single RF antenna below or around the sides of a homecage17,19. Due to electromagnetic dissipation away from the RF source, wireless coverage can be as low as 30% in a homecage20 and worse in larger behavior boxes. Previously, these limitations were circumvented by increasing RF power, but this results in undesired RF energy to animal tissues and increased heat generation. A recent study utilized an RF multiplexer that rapidly toggled power between two antennas to increase coverage21. However, this approach requires careful operation and validation to avoid electromagnetic interference between the antennas, and limits the use of a multiplexer for powering multiple cages. Here, we introduce a simple dual-coil-antenna system for increasing wireless coverage22. It consists of a top antenna coil that is connected to an RF generator and an unconnected antenna coil below the cage that passively attracts RF signals towards the animal subject and cage bottom. Three-dimensional electromagnetic modeling suggested that the dual-coil-antenna system could enable continuous operation throughout a location of interest (Fig. 2e; top and Supplementary Fig. 10). This was confirmed with light-power-output measurements of wireless devices at five representative positions and heights from the cage bottom, which demonstrated robust device activation throughout the volume of a cage (Supplementary Fig. 11). Furthermore, the dual-coil-antenna system eliminated the dependence of transmitted power on the relative orientation angle between the transmission antenna and the device (Supplementary Fig. 12). Comparison studies further indicated that the proposed antenna system outperforms other existing systems, offering virtually complete wireless coverage in a homecage (Fig. 2e (bottom) and Supplementary Fig. 13).

Scalable, multimodal device operation

Multimodal device operation is another strategy for increasing the efficiency and throughput of wireless optogenetic studies. Targeting multiple organs with a single device could enable multiorgan analysis in the same animal or even be used to examine organ-to-organ interactions. Realization of multimodal tools requires an actuation mechanism that can remotely manage channel selection. Previous efforts utilized higher operating frequencies, micro-controller chip, or Bluetooth kits for actuating separate channels, but these approaches require increased RF power (tens of mW) for operation and render them energy-hungry devices10,21. Here, we use a reed switch in the device that responds to the pattern of externally applied electromagnetic RF pulses. In our example, a pulse width longer than 100 ms triggers the transition from a green µLED to a blue µLED located on a separate tether (Fig. 2f and Supplementary Movie 3). The actuation threshold can be adjusted by pairing different capacitors and resistors with the reed switch to prevent unwanted activation or deactivation, and could theoretically be tuned for switching between more than two channels23,24; the circuit diagram is shown in Supplementary Fig. 14. Importantly, this strategy only requires 10 µW for channel selection, which is 100-fold less power than other approaches10,21,25. When combined with the dual-coil antenna and multiplex coupling/decoupling, the proposed optoelectronic system enables robust, ultra-efficient wireless powering of optogenetic devices in multiple organs and multiple cages with independent and simultaneous control. Time slots allocated for each cage and threshold pulse for channel selection are tunable, suggesting many scenarios of multiplexing and multimodal operation. For example, we set the threshold pulse for activation/deactivation of channels to 100 ms and allocate 250 ms for each cage. This provides enough time for an implant in each cage to switch its channel (from Ch1 to Ch2 or vice versa). It requires only 2 s (8 × 0.25 s) for the switching operation of implants in cages. Next, the TX system can adjust time slots, depending on stimulation conditions.

Optogenetic manipulation of gastric vagal sensory endings

To determine the utility of the optoelectronic system, we investigated the role of stomach vagal afferent endings in feeding behavior. We began by analyzing the µLED light spread and identified RF powering parameters needed for organ specificity. As expected, securing the µLED inside the stomach significantly restricts light spread, in contrast to surface affixation, which results in light back-scatter intensities well above the threshold for opsin activation26 (Fig. 3a and Supplementary Fig. 15). We examined whether implantation of the stomach device is well tolerated by showing that ad libitum food intake of mice implanted with the device was the same as sham-operated mice (Fig. 3b). These results indicate that our wireless device should allow precise optogenetic manipulations in awake, behaving mice.

Fig. 3: Optogenetic targeting of Calca+ vagal afferents in the stomach.
figure 3

a Light intensity measurements comparing LED implantation inside vs. outside the stomach (n = 5, p < 0.01), with varying RF powers (p < 0.001). Dashed horizontal lines indicate light intensity needed for 10 and 50% maximal activation of channelrhodopsin2. b Comparison of total food intake, number of meals, and meal size in mice implanted with LED device (n = 7) or sham operated (n = 6) (p = 0.71). c CalcaCre transgenic mice received nodose ganglion injection of AAV9-DIO-ChR2:tdTomato. Images show fluorescence in situ hybridization of tdTomato and Calca mRNA, demonstrating the cell-type specificity of transgenic/viral approach; scale bars 25 µm. d tdTomato fluorescence labeling of central Calca+ vagal afferent endings in the nucleus of the solitary tract (NTS); scale bar 25 µm. e Fluorescence labeling of peripheral Calca+ vagal afferent endings in the stomach mucosal layer; scale bar 50 µm. Behavioral experimental results are from one cohort of animals. Bar graphs are mean ± SEM. Statistical comparisons were made using two-way repeated-measures ANOVA, Tukey’s post hoc; ***p < 0.001.

A recent study identified genetically distinct vagal afferent neurons in the nodose ganglion that innervate the stomach and express either Calca, Sst, Gpr65, or Glp1r genes27. In contrast to Sst and Gpr65, which exhibit either mechanosensitive morphological endings in muscle layers (Glp1r) or chemosensitive endings in the mucosal layer (Sst and Gpr65), Calca+ neurons form spatially restricted chemosensitive mucosal endings in the corpus vs. mechanosensitive intramuscular arrays in the stomach antrum27. Identification of a role for stomach chemosensation in appetite control has been elusive28; therefore, we used our wireless device to selectively activate Calca+ vagal afferent chemosensitive endings in the corpus region of the stomach.

To gain cell-type specificity, AAV9 was injected into the nodose ganglion of CalcaCre transgenic mice to introduce Cre-dependent ChR2:tdTomato opsin expression or a control group with just tdTomato fluorescent reporter (Fig. 3c). Precise anatomical specificity was achieved by implanting the µLED into the fundus, immediately adjacent to the corpus (Fig. 3d). While Calca+ vagal afferents do not innervate the fundus, we implanted the device away from the antrum to avoid activation of mechanosensitive fibers (Fig. 3e). Several weeks after recovering from device implantation, mice were fasted overnight and refed the following morning. Compared to no stimulation (RF antenna off), optogenetic activation produced robust suppression of food intake during refeeding, with greater stimulation frequencies almost completely suppressing intake (Fig. 4a, b and Supplementary Movie 4 (left)). Importantly, activation of the device in the control group without ChR2 did not alter feeding behavior, indicating that RF signals and activation of the device in itself do not influence feeding (Fig. 4c and Supplementary Movie 4 (right)). To further establish that the appetite suppression was due to activation of vagal afferent endings in the stomach, we compared these results to separate cohorts of mice implanted with six non-attached µLEDs in the abdomen. Although optogenetic stimulation of vagal afferents in this manner suppressed feeding, the effect was not as robust despite stimulating with six µLEDs rather than a single µLED directly implanted inside the stomach (Supplementary Fig. 16). Furthermore, the non-anchored LED approach required increased operating power to compensate for light dissipation, resulting in greater heat generation and potential tissue damage29,30. Thus, the gastric optogenetic implant enables more robust optogenetic activation using less wireless power.

Fig. 4: Activation of Calca+ stomach vagal afferents suppress appetite via negative valence mechanism.
figure 4

a CalcaCre transgenic mice received a left nodose ganglion injection of AAV9-DIO-ChR2:tdTomato or AAV9-DIO-tdTomato control virus. The LED was implanted in the stomach corpus-function junction. b Frequency-dependent suppression of food intake in the ChR2:tdTomato group (n = 8). c The tdTomato control group did not suppress food intake during photostimulation (n = 4) (p = 0.06). d Illustration of real-time place-preference (RTPP) box. The RF antenna powered the device only in the right chamber. e Activation of LED device (20 Hz light pulses) did not induce a place preference nor avoidance in both ChR2 and tdTomato groups (n = 7 per group) (p = 0.31). f Representative traces for RTPP assay. g Illustration of a large open-field box; the antenna delivered wireless power throughout the entire arena (20 Hz light pulses). h Photoactivation of Calca+ gastric vagal afferents decreased time spent in center (n = 7 per group). i Representative traces from open-field test. j Mice were exposed to a novel sucrose solution on Day 1 followed by optogenetic activation of vagal sensory fibers (20 Hz). On Day 5, mice were water-restricted overnight and then given simultaneous access to a bottle of sucrose and a bottle of water. The graph is the sucrose preference score (ChR2, n = 7; tdT, n = 5). Experimental results are from one cohort of animals. Bar graphs are mean ± SEM. Statistical comparisons were made using two-way repeated-measures ANOVA, Tukey’s post hoc, except for (h) and (j), which were two-tailed t test; **p < 0.01, ***p < 0.001.

Appetite suppression can be associated with positive valence, potentially due to the removal of aversive hunger signals31, or aversion in response to harmful stimuli, such as uncomfortable gastric distension32 or food poisoning33. To investigate the affective mechanisms by which Calca+ gastric vagal afferent neurons might suppress appetite, we constructed oversized dual-coil antennas for robust optogenetic activation in various behavior boxes (Fig. 4d, g and Supplementary Fig. 17). In one assay, mice were placed in a two-chamber box with RF power only in one chamber to determine whether mice form an aversion or preference for the optogenetic stimulation chamber. Surprisingly, we did not observe differences in place preference or avoidance (Fig. 4e, f), similar to optogenetic stimulation of other vagal afferent cell types that innervate the gastrointestinal tract27. Conversely, an open-field assay demonstrated that optogenetic stimulation reduced the time mice spent in the center, indicative of anxiety-like behavior and suggesting that activation of Calca+ gastric vagal afferent fibers might be aversive (Fig. 4h, i). We analyzed the locomotor activity from open-field and place-preference tests, which revealed decreased locomotion during optogenetic activation of Calca+ gastric vagal afferent (Supplementary Fig. 18); presumably, the mice feel aversion and have decreased motivation. In addition, gastrointestinal signals are closely associated with taste-sensory signals34, suggesting that a learned food preference/aversion assay might be most indicative of the mechanism underlying appetite suppression. To test this hypothesis, mice were habituated to overnight water restriction for several days and then given access to 5% sucrose solution, followed by optogenetic stimulation of Calca+ vagal afferent fibers for 4 h. Three days later, mice were offered the choice of water or 5% sucrose. This two-bottle preference test revealed that activation of stomach Calca+ vagal afferents conditioned mice to avoid the sucrose solution (Fig. 4j). This suggests that appetite suppression occurs via a negative valence mechanism that alters taste preferences. These results identified a role for stomach mucosal Calca+ vagal afferents in appetite suppression and revealed a mechanism by which appetite suppression occurs.

Discussion

While many classical studies have established an important role for visceral signals in controlling behavior35,36, these surgical- and chemical-denervation experiments lacked organ specificity and did not reveal the identity of sensory neurons that can serve diverging functions. Here, we developed wireless µLED devices that permit organ-specific, optogenetic manipulations, and an ultra-efficient wireless telemetry system for powering multiple cages. The miniaturized wireless device enabled precise optogenetic stimulation of genetically defined vagal afferents innervating the mouse stomach, revealing a function for Calca+ mucosal sensory endings in suppressing food intake via a negative valence mechanism. Critically, the pre-curved sandwich construction significantly extended the lifespan of the µLED device and allowed for testing various stimulation parameters and behavior tests within the same subjects. We envision that the current device could be used to optogenetically manipulate neural circuits throughout the gastrointestinal tract and other hollow organs, such as the bladder with little or no modifications.

Prior methods for optogenetic activation of vagal afferents in awake mice have either lacked organ specificity27 or involved gut injections of a retrogradely transported opsin virus and fiber-optic implantation in the hindbrain where vagal afferents terminate4. While the latter provides organ specificity, retrograde viruses can be limited by tropism and incomplete infection of certain cell types. Moreover, vagal afferents express neuropeptides in their peripheral endings which are hypothesized to be released in the gut to exert efferent functions7. In other systems, such as for somatosensation, the peripheral release of neuropeptides by afferent fibers can contribute to behaviors by sensitizing other afferent subtypes to ongoing stimuli37. Finally, fiber optics cannot be used for optogenetically manipulating the enteric nervous system nor splanchnic sensory afferents, which synapse in the spinal cord. Investigating the function of these neural circuits and hypotheses, therefore, requires peripheral optogenetic stimulation that is now possible with the proposed wireless gastric optogenetic implant. Multimodal features could further enable the investigation of peripheral interactions by using different colored µLEDs to activate corresponding color-sensitive opsins expressed by separate neural substrates or multiple organs simultaneously/independently.

In addition to extending optogenetic functionality to the peripheral nervous system, we introduced advancements in wireless telemetry that generally improve the scalability and usability of optogenetics. The dual-coil-antenna system, which enables reliable and complete wireless coverage, is easily constructed using inexpensive copper wire secured onto cardboard or plastic backing. The multiplexing approach further allows for the testing of large experimental cohorts, which was particularly important for our studies because of the extended duration of feeding behavior tests. This system can be set up in under an hour, is simple to operate, and dramatically decreases the cost and time required for conducting optogenetic experiments. Furthermore, the wireless telemetry system has broad applicability for powering optogenetic devices in the periphery, brain9,38, or other wireless devices, such as those that measure bioelectrical signals10.

Future studies may take advantage of these enabled wireless optoelectronic features to chronically activate neural circuits for days, weeks, or even months. Because adaptations can occur with sustained activation of a neural pathway, such experiments are important for investigating the persistence of long-term, physiological effects, including weight loss. This can enable experiments that determine whether appetite suppression induced by gastric vagal afferent activation is attenuated in obese mice and whether chronic activation of vagal afferent endings in the stomach can reverse obesity. Identification of viscerosensory pathways that can either suppress or stimulate appetite will have direct clinical importance for potentially developing novel therapeutic targets for treating appetite disorders.

Methods

Device fabrication

The process began with flexible copper/polyimide (Cu/PI) bilayer films (thickness; 12 µm/18 µm, AC181200RY, DupontTM Pyralux®) mounted onto a glass slide (dimensions, 5.08 cm × 7.62 cm). Then, we deposited 2.5-µm thickness of photoresistor on the Cu/PI substrate (AZ 1518, AZ®, recipe; spin coated at 3000 r.p.m. for 30 s), and used UV photo-lithography to define patterns for pads and interconnections (EVG610, EV Group, recipe; UV intensity for 200 mJ cm−2). This was followed by immersion in developer solution (AZ Developer 1:1, AZ®) for 30 s and rinses in distilled water for 10 s. Immersion in copper etchant (LOT: Z03E099, Alfa AesarTM) for 7 min and rinses with acetone, methanol, isopropanol, and distilled water for 1 min yielded Cu interconnections and pads on the flexible substrate (Fig. 1c; step 1). After samples dry, chip components were mounted, including a µLED, passive components, and IC components using a soldering machine. An additional PI/Cu layer (18 µm/12 µm thick) with the bottom chip-mounted Cu/PI substrate formed a sandwiched structure (PI/Cu/Cu/PI) (Fig. 1c; step 2). For encapsulations, we applied a small amount of PDMS (SylgardTM 184 Silicone Elastomer Kit, Dow®; 10:1 mix ratio) using a pipette while a clamp held the body of a sample to form a thin, pre-curved, sandwiched structure (Fig. 1c; step 3). Then, we encapsulated the body of a sample with PDMS by a dip-coating process (500 µm thick). Samples were cured in a vacuum oven at 100 °C for 1 h. These procedures yield a soft, low-power, wireless gastric optogenetic implant with a pre-curved, sandwiched tether (Fig. 1c; step 4). Detailed information on device layouts, IC components, and procedures for fabrication are found in Supplementary Fig. 1 and Supplementary Table 1.

Finite-element method analysis

For numerical electromagnetic simulations of the proposed device, we used a finite-element method analysis tool (Ansys Electromagnetics Suite 17-HFSS, Ansys®) with Cole-Cole dielectric relaxation model where characteristics of biological tissues were described as a function of frequency. Organ systems and tissues of a mouse were modeled to one million meshes for numerical simulations, and antenna coils made of copper stripes or wires were modeled to materials with finite conductivity, 58 MS s−1. For three-dimension modeling of the mechanics for the devices, we used a commercial finite-element method analysis tool (Abaqus/CAE 2018, Dassault Systems) to investigate strain effects on the pre-curved and post-curved structures. The following parameters were used for simulations: thickness 500/18/12/12/18/500 µm (PDMS/PI/Cu/Cu/PI/PDMS) for the pre-curved structure and 510/12/18/510 µm (PDMS/Cu/PI/PDMS) for the post-curved structure; elastic properties Young’s modulus (MPa)/Poisson’s ratio: 1/0.49 for PDMS, 119000/0.34 for Cu, and 2500/0.34 for PI. Cu/PI layer was modeled as a composite shell element (S4R). PDMS was modeled as a solid hexahedron element (C3D8R) in the pre-curved structure and as a shell element (S4R) in the post-curved one. The mechanical simulation results are found in Supplementary Fig. 2.

Mechanical, optical, and electrical measurements

We used a gauge-force machine (ESM303 Forced Test Stand, MARK-10) to perform device lifetime cycling tests with a significant load extended over a period of time (>200 kc) for the pre-curved, post-curved, and pre-curved structures with three different curvatures (0.72, 1.15, and 2.87 mm) of a tether. The experiments involved the application of strain in three different directions: (1) the x-direction, (2) the y-direction, and (3) the z-direction, respectively. After completion of each 1000 cycles, we immersed a wireless device in 10% phosphate-buffered saline (PBS) solution for 10 min and measured light intensity using a light meter (LT300, Extech). This test was repeated until a device stopped functioning (Fig. 1e, f and Supplementary Fig. 4). We also performed accelerated life testing where a device was immersed in 10% PBS solution and light intensity was monitored as a function of time at various temperatures (25, 60, and 90 °C) (Supplementary Fig. 5). For thermal assessments of wireless devices, we used an infrared camera (VarioCAM HDx head 600, InfraTech). Light intensity was fixed at an optical intensity of 10 mW mm−2, which is enough to activate light-sensitive proteins, and the camera measured variations in temperature when devices were operated with duty cycles of 20, 40, 60, 80, and 100% (Supplementary Fig. 6).

Antenna-coil fabrication and wireless, power-control system

We used 8-gauge bare Cu wire for the bottom antenna coil and Cu stripes (0.635 mm thick by 2.54 cm wide) for the top antenna coil. The bottom coil was placed under a cage while the top coil was situated 8 cm above the cage bottom. Impedance matching using Network Analyzer (ENA Series E5063A, Keysight) with a discrete capacitor component yielded two antenna coils, each of which resonates at 13.56 MHz (the top coil) and 15 MHz (the bottom coil), respectively; these different frequencies offer broad bandwidth and stable coverage. Wireless power-control systems consisted of an RF-power supply (ID ISC.LRM2500-A, FEIG Electronics), matching board (ID ISC.DAT-A, FEIG Electronics), RF multiplexer (ID ISC.ANT.MUX.M8, FEIG Electronics), controller (nRF52832 Development Kit, Nordic semiconductor), and decoupling multiplexer. The controller was programmed a custom C code based on C code libraries from Nordic (nRF5_SDK_13.0.0_04a0bfd) by Keil uVision 5 IDE (µVision V5.23.0.0). The circuit diagram, device layout, and information on decoupling multiplexer are found in Supplementary Fig. 9. Dimensions of a representable coil antenna, capacitance, and inductance for different sizes of experimental assays including homecages are found in Supplementary Table 2.

Measurements of wireless coverage

We implanted a wireless device over the skull, under the skin of a mouse, and recorded their behaviors using three cameras (C615, Logitech). A red-colored µLED was embedded in an implanted device to serve as a signal that can be easily detected by cameras over a cage and the wireless TX system transmitted RF signals at 1 W. One camera was positioned above a cage and two cameras recorded from left and right sides. They recorded behaviors of an animal in a cage for 2 min and we extracted images from the recordings and analyzed them frame by frame to determine whether an image had captured wireless operation of a device (red µLED). Next, we counted the number of frames missing wireless operation. For the purpose of visual demonstration of wireless coverage, we reconstructed 3D continuous traces of a red µLED from extracted images using custom scripts in Python (version 3.7.3–64 bit, Spyder 3.3.6 IDE). We repeated the procedures described above for other wireless antenna technologies (Supplementary Fig. 10). For validations of wireless power TX systems, we used an electromagnetic probe (TBPS01-TBWA2/40 dB, Tekbox) to measure the output power at five representative positions (A, B, C, D, and E) and various heights from the bottom of the enclosure as a function of the distance and angle (Supplementary Figs. 11 and 12).

Mice

CalcaCre:GFP homozygous mice (C57Bl/6 background; Jackson laboratory Calcatm1.1(cre/EGFP)Rpa) were bred with C57BI/6 mice to generate CalcaCre:GFP/+ heterozygous mice used in experiments39. Following surgery, mice were singly housed with ad libitum access to standard chow diet (LabDiet 5053) in temperature- and humidity-controlled facilities with 12 h light/dark cycles. Both male and female mice were used for behavioral experiments. All animal care and experimental procedures were approved by the Institutional Animal Care and Use Committee at the University of Washington.

Organ-specific, wireless, gastric optogenetic device implantation

Under surgical anesthesia (isoflurane, 1–2% inhalation), the animal’s ventral side was shaved, sterilized with three alternating scrubs of betadine and alcohol, and the surgical field was restricted with sterile drapes. With the animal on its back, a 2 cm skin incision was made along the abdominal midline from the xiphoid cartilage extending to the mid-abdomen, and a second cut into the abdominal wall exposed the stomach for device implantation. Ringed forceps were used to gently grasp the fore-stomach and pull it out of the abdominal cavity onto gauze soaked with sterile saline. Fine-tipped Dupont forceps were then used to puncture the stomach fundus and thread the µLED tether in and out of the stomach. With the µLED inside the stomach, the tether was secured in place with purse-string sutures (5-0 PGA). The device harvester was then placed in the abdominal cavity and the stomach was placed back into its normal orientation. The abdominal wall was closed with interrupted stitches using absorbable suture (5-0 PGA), and the skin with non-absorbable suture (6-0 silk). Mice received analgesics during the surgery (ketoprofen, 5 mg kg−1) and daily post-operative care (provided with hydrating gel, monitor food intake, and body weight). For multimodal device implantation, an incision was made in the abdominal cavity and the device was implanted with the blue LED positioned towards the thoracic cavity and the green LED towards the abdominal cavity. After recovery from surgery, all animals received daily post-operative care and monitoring.

Meal-pattern analysis

To examine whether mice tolerate stomach device implantation, one group of mice was implanted with the device, whereas another group underwent a sham surgery where the abdomen was opened near the stomach, but it was not punctured nor implanted with a device. Two weeks after the surgeries, mice were placed in food-monitoring homecages (BioDAQ, v. 2.2). Feeding records were analyzed using BioDAQ Viewer (software v. 2.2.01). A feeding bout (≥0.01 g) was defined as a meal if ≥0.06 g of food was ingested and if it was separated from another meal by ≥5 min.

Nodose ganglion injection

With the mouse under anesthesia (isoflurane, 1–2% inhalation), a 1–2-cm-long skin incision was made from the left clavicle going upwards to the animal’s jaw. The left vagus nerve was exposed by separating the platysma, sternohyoideus, and omohyoideus muscles using blunt dissection. After visualization of the ganglion, the virus (200 nl) was injected with a glass micropipette attached to a Nanoject II. The experimental group received AAV9-DIO-ChR2:tdTomato and control groups AAV9-DIO-tdTomato. The skin was then closed with interrupted stitches (6-0 silk suture). After the experiments, mice were anesthetized (Beuthanasia, 320 mg kg−1 delivered intraperitoneally) and intracardially perfused with PBS, followed by 4% paraformaldehyde. Brains, nodose ganglion, and stomach tissues were then extracted, post fixed in 4% paraformaldehyde overnight, and cryoprotected in PBS containing 30% sucrose until the tissues sunk in the sucrose solution. Coronal cryostat sections for the brain, nodose, and stomach tissue were collected (30, 20, and 10 µm thick), directly mounted onto microscope slides, and coverslipped using DAPI Fluoromount-G mounting medium (SouthernBiotech). Two mice injected with AAV9-DIO-ChR2:tdTomato were excluded from behavioral analysis due to little (2–3 neurons infected per section) or no virus infection as determined by visualizing the fluorescent reporter, tdTomato.

In situ hybridization

Single-molecule fluorescence in situ hybridization was performed using an RNAscope Fluorescent Multiplex Kit (Advanced Cell Diagnostics). C1 and C3 DNA oligonucleotide probes were designed for Calca and tdTomato. Nodose sections were fixed in 4% paraformaldehyde for 15 min and then washed in 50, 70, 100, and 100% ethanol for 5 min each. Slides were dried for 5 min. Proteins were digested using protease solution (pretreatment solution 3) for 60–90 s. Immediately afterward, slides were washed twice in PBS. In parallel, C1 and C2 probes were heated in a 40 °C water bath for 10 min. Probes were applied to the slides, which were coverslipped and placed in a 40 °C humidified incubator for 3 h. Slides were rinsed twice in RNAscope wash buffer and then underwent the colorimetric reaction steps according to the standard kit protocol. After the final wash buffer, slides were immediately coverslipped using DAPI (4′,6-diamidino-2-phenylindole) Fluoromount-G mounting medium. Images were captured using a laser-scanning confocal (FV1200, Olympus) and epifluorescent (Eclipse E600, Nikon) microscopes.

Measurement of µLED light spread

Light intensity measurements (PM100D, Thorlabs) were acquired from fresh stomach tissue preparations ex vivo. Measurements were recorded after implanting the LED inside the stomach or after suturing the tether to the stomach surface with the LED directed towards the organ; the light sensor (S130C, Thorlabs) was encased in saran wrap and placed directly over the LED tether housing. We tested 2, 4, 6, 8, 10, and 12 W antenna power outputs. In a separate experiment, with LED not attached to a stomach, light dissipation was measured by placing the sensor 0, 0.5, and 1 cm away from the front- or backside of the LED housing.

Fasting and refeeding experiments

Mice were food restricted overnight (16 h) and refed the following morning. Food intake was manually measured 1, 2, and 3 h after refeeding. The same animals underwent multiple fasting–refeeding tests to examine different optogenetic stimulation parameters: no stimulation (RF antenna off), 10 Hz, and 20 Hz optogenetic stimulation (5 ms pulse width; RF power 4 W). Experiments were conducted 5 days apart.

Real-time, place-preference (RTTP) assay

Mice were placed in an RTPP box consisting of two chambers (20 cm × 18 cm) and a small transition area. Antennas were installed in both chambers, but only one chamber was connected to an RF generator to continuously deliver RF power (20 Hz, 5 ms pulse width, 4 W). The time spent in each chamber (20 min trial) was analyzed using video-tracking software (EthoVision XT 10, Noldus).

Open-field test

Mice were placed in the center of a 40 cm × 40 cm square open-field arena with non-transparent white Plexiglas. The total distance moved and time in the center (20 cm × 20 cm imaginary square), during the 10 min trial, were analyzed with video-tracking software with EthoVision. An RF antenna provided wireless power (20 Hz, 5 ms pulse width, 4 W) throughout the entire behavior box.

Two-bottle flavor preference test

Mice were accustomed to drinking water from two test tubes that replaced their normal water bottles. After acclimation, mice were water-deprived overnight, and the following morning received 30 min access to a novel 5% sucrose solution; immediately following sucrose consumption, mice received 4 h of optogenetic stimulation (20 Hz, 5 ms pulse width, 4 W). Three days later, mice were water-deprived overnight, and the following morning received 30 min access to separate test tubes containing either water or 5% sucrose solution. The intake of both solutions was measured and presented as a preference ratio (5% sucrose intake/total intake of sucrose and water solutions).

Statistics

Data were analyzed using Prism 5.0 (GraphPad software). Sample sizes were estimated based on prior experience and expected variability in feeding behavior27. We excluded an animal from data analysis if post hoc histological analysis showed no viral transduction as indicated by an absence of tdTomato fluorescence. For graphs comparing two experimental conditions, we used unpaired two-tailed Student’s t test. We analyzed data sets (multiple treatments and time-points) with repeated-measures two-way analysis of variance tests (time repeated factor) and Tukey’s post hoc tests. All data sets were conducted using Shapiro–Wilk normality test, and all passed the normality tests.

Reporting summary

Further information on research design is available in the Nature Research Reporting Summary linked to this article.